Firmware Installation Procedure (Install Fw); Id Writing Procedure (Write Id); Latest Firmware Disk (As Of May 1998) - Agilent Technologies E4915A Service Manual

Crystal impedance / lcr meter
Table of Contents

Advertisement

Firmware Installation Procedure (INSTALL FW)

1. Prepare a rmware disk. The Agilent Part Number for the latest rmware disk (REV.2.10) as
of May 1998 is E4915-18006.
2. Setup the PC environment which is used for adjustments and performance tests of the
E4915A/E4916A. (see Adjustments and Performance Tests )
3. Connect the E4915A/E4916A to the PC through the GPIB cable.
4. Start the Adjustment/Performance Test Program.
5. Select \INSTALL FW" in the list box on the program display and click
6. Follow the instructions of the program.
Note
New rmware installation is required after A1 Main assembly is replaced.

ID Writing Procedure (WRITE ID)

1. Setup the PC environment which is used for adjustments and performance tests of the
E4915A/E4916A. (see Adjustments and Performance Tests )
2. Connect the E4915A/E4916A to the PC through the GPIB cable.
3. Start the Adjustment/Performance Test Program.
4. Select \WRITE ID" in the list box on the program display and click
5. Follow the instructions of the program.
Note
a. If the Model, Serial No., and Option information in the \Check ID DATA
(Write)" eld in the \WRITE ID DATA" window is incorrect, you must click
NNNNNNNNNNNNNN
EXIT , click
Model" window to exit the program, and then click
program.
b. The installation status of the E4916A Option 001 (Add Impedance Probe)
is not saved into the E4916A's internal memory. This means \001" is not
appeared on the \ID DATA SET AS FOLLOWS :" eld in the \CHECK ID
DATA" window.
c. ID writing must be performed after A1 Main assembly is replaced.
Table 7-2. Latest Firmware Disk (As of May 1998)
Agilent Part
Qty.
Number
E4915-18006
1
NNNNNNNNNNNNNNNNNNNN
Cancel in the \List Box" window, click
Description
Firmware Disk Rev. 2.10
NNNNNNNN
OK .
NNNNNNNN
OK .
NNNNNNNNNNNNNN
EXIT in the \Select
NNNNNNNNNNNNNNNNN
Start to restart the
Post Repair Procedures 7-3