Exchanging a Flow Cell

When
If an application needs a different type of flow cell or the flow cell is defective (leaky).
Tools required
Two 1/4 inch wrenches for capillary connections
Parts required
#
1
1
Preparations
Turn off the flow.
DO NOT install the inlet capillary to the outlet connection of the flow cell. This will result in
N O T E
poor performance.
In case the flow cell is not used for some time (stored), then flush the flow cell with
N O T E
iso-propanol and close the cell with Plug-Screw 10-32- Fitting (p/n 0100-1259).
1
Press the release buttons and remove the front cover for
access to the flow cell area.
Agilent 1260 FLD User Manual
p/n
Description
G1321-60005
Standard flow cell, 8 µL, 20 bar
G1321-60007
FLD Cuvette Kit, 8 µL, 20 bar
2
Disconnect the capillaries from the flow cell.
Maintenance

Exchanging a Flow Cell

9