Exchanging a Flow Cell

When required
Tools required
Parts required
DO NOT install the inlet capillary to the outlet connection of the flow cell. This will result in
N O T E
poor performance.
Preparations for this procedure:
Turn off the flow.
1200 Series FLD User Manual
If an application needs a different type of flow cell or the flow cell is defective
(leaky).
Two 1/4 inch wrenches for capillary connections
Standard flow cell, 8 µl, 20 bar, G1321-60005
Cuvette for off-line measurements, 8 µl, 20 bar, G1321-60007, refer to
use the Cuvette"
on page 116 for more information on usage.
1 Press the release buttons and remove the front cover for
access to the flow cell area.
Maintenance
8
"How to
113