Urodynamic interface
•
Set Out H-Centre to 79 (press the + or the - key).
•
With the Menu key, select Exit and use the + to return to the Advanced menu.
•
Press the Menu key until Manual Set appears.
•
Press the + key (the sub-menu is displayed).
- Select VGA Left and set 18, until only the X-ray image is visible (press the + or the -
key, default setting is 9).
•
Use the Menu key to select VGA Store for a save operation and press the + key until an
acoustic signal is emitted.
•
To save all settings, press the Menu key and the + key at the same time..
•
Switch the UROSKOP Access off and back on again.
Cutting off the menu bar (FLUOROSPOT Compact)
NOTE
•
Press the Menu key until Adjust: Screen size appears.
•
Press the + key.
- The sub-menu is displayed.
•
Set Out H-Centre to 65 (as setting for PAL; setting for NTSC is 66) (press the + or
the - key).
•
With the Menu key, select Exit and use the + key to return to the Advanced menu.
•
Press the Menu key until Manual Set appears.
•
Press the + key (the sub-menu is displayed).
- Select VGA Left and set 1, until only the X-ray image is visible (press the + or the -
key).
•
Use the Menu key to select VGA Store for a save operation and press the + key until an
acoustic signal is emitted.
•
To save all settings, press the Menu key and the + key at the same time..
•
Switch the UROSKOP Access off and back on again.
Scan converter parameter table (ASPIA imaging system)
Tab. 1
Scan converter parameter table (ASPIA )
Parameters
Menu"Screen Size"
Out H-Centre
Out H-Width
Out V-Centre
Siemens AG
Medical Solutions
This subsection refers exclusively to UROSKOP-Access systems
with the FLC imaging system.
Default value (PAL)
1
79
93
160
SPL5-330.814.01.05.02
08.05
Default value (NTSC)
79
94
132
Page 13 of 18
CS PS 24
13
0
0
UROSKOP ACCESS