PC Board Replacement
4-

Component N 11

3PC Board Replacement
NOTE

Board D1, (new type) host computer or when replacing the battery

78 28 841 J 1042,
30 82 138
Follow items A + B + D when replacing board D1.
Follow items A + C + D when replacing the battery.
A
If possible, perform a backup of all current parameters and data.
(Refer to Operation of the Service Software RXL2-120.113.01. Chap. 3-3)
Read out the counter values for
shock counter generator
shock counter spark-gap
shock counter US probe (starting with Rev. 05)
and document them in the operating log (Logbook, Reg. 8).
Remove board D1 from electronics rack N11.
B
Change EMC shielding plate, if present, from old to new board.
Set or check the switches on the new board.
Switch settings, see wiring diagram excerpt in the Appendix to this manual.
C
Measure the battery voltage; U
If the battery voltage is < 2.7 V, replace the battery.
Desolder the old battery and replace it with a new one.
D
Measure the battery voltage; U
If the battery voltage is < 2.7 V, replace the battery.
Insert board D1.
After switching on the system:
Siemens AG
Medical Solutions
Refer to wiring diagram J1042-3A for the rated current values of
the different fuses listed in this chapter.
RXL2-120.841.02.01.02
09.05
refer to wiring diagram J1042-21-4
battery 3.0 V
> 2.7 V is normal.
batt
> 2.7 V is normal.
batt
Page 21 of 72
CS PS 24
shock counter shockwave
shock counter total