Replacing the collimator lamp

Equipment required: Lux meter
Remove the cover (1) by removing the screws (4) marked "A" (Fig. 1). Turn the collimator
around so that the lamp housing is in front of the cover opening.
Remove the protective plate (2) covering the collimator lamp by loosening, not removing,
the screws marked "B" by one turn. Do not adjust screws "C", "D", or "E" (Fig. 2).
Remove the defective lamp and replace it with a new lamp (Fig. 3). Do not touch the
bulb with bare fingers. (See "list of spare parts".)
Reinstall the protective plate. Tighten the screws marked "B".