Replacing The Lamp For The Localizer - Siemens POLYMOBIL Plus Service Instructions Manual

Hide thumbs Also See for POLYMOBIL Plus:
Table of Contents

Advertisement

6 - 4
7
8
Fig. 6

Replacing the lamp for the localizer

The replacement light localizer lamp is located below the cassette module.
Turn both knobs (8/Fig.6) to the end position and remove them.
Remove the front panel after removing both screws (7/Fig.6) .
Caution: The front panel is connected to the collimator via a cable (9/Fig.7).
Remove both screws (10/Fig.7) and take off the metal bracket for the temperature
switch).
Take out the defective lamp (11/Fig. 7) and replace it with a new lamp.
CAUTION
Attach the bracket and the front panel.
Adjust both knobs and attach them.
Check the function of the light localizer lamp.
Check the considence of light and radiation fields (see "Considence of light and radiation
fields" in chapter 5).
POLYMOBIL Plus
Replacing important components
Do not touch the glass envelope with your fingers.
SPR8-125.061.02
Rev. 01
Fig. 7
Page 4 of 8
01.00
TD SD 24
10
9
11
Siemens AG
Medical Engineering