2.4 Electrical Connections

2.4.1 Foxy 200 Control Cable

Locate the serial cable, P/N 480-7996-00, in the accessory kit.
Attach this cable's 9-pin connector to the Foxy 200 Control
Port on the rear panel of the Companion XL. Attach the other
end to the 25-pin connector labeled "SERIAL" on the back of
the Foxy 200.
If additional Foxy 200 Fraction Collectors are to be used with
the system, connect the additional fraction collectors using the
supplied "Y" cables.

2.4.2 Connect Power

Mains power must meet the voltage, frequency, and amperage
requirements listed on the instrument's serial number label.
Use the supplied IEC power cords to connect the
Companion XL and Foxy 200 to mains power.

2.5 Connect Solvent Tubing

Attach the Solvent A and B lines to the fittings on the back of
the Companion XL. Place the weighted ends of the solvent
tubing in their respective solvent containers.
The solvent ports at the back of the Companion XL can be
rotated so that tubing can be routed easily. To rotate the
fitting, loosen the Swagelok
the desired position, then re-tighten the nut.
Note
To prevent damage or premature wear to the pump and internal
valves, clean solvent should be used. The solvent should not
contain any dissolved solids.
Risk of fire or equipment damage. Failure to connect Foxy 200
tubing may allow organic solvents to pool in unsafe areas,
creating a potential for dangerous levels of flammable vapors.