Troubleshooting
the System
Troubleshooting
Patient Location
Error Message
Possible Cause
Action
32002 E:INVALID-MESSAGE
Data from SmartLinkTM is
Contact Fisher-Berkeley service
invalid
representative
32003 E-GARBLED-MESSAGE
Data from SmartLinkTM is
Contact Fisher-Berkeley service
invalid
representative
32004 E-BAD-LRC
Data from Sma.rtLinkTM is
Contact Fisher-Berkeley service
invalid
representative
32005 E-RS232-OPEN-FAILED
A problem exists in the
Re-seat the RS-232 board. Replace
RS232/422 interface
the board if the problem does not go
away.
32006
A problem exists in the
E-RS232-CONFIG-FAILED
RS232/422 interface
Re-seat
the
RS-232 board. Replace
the board if the problem does not go
away.
32007
E-RET-PENDING-MSG-BUF
A problem exists in the
RS2321422 interface
Re-seat the RS-232 board. Replace
the board if the problem does not go
away.
32008 E-INVALID-TX
A problem exists in the
RS232/422 interface
Re-seat the RS-232 board. Replace
the board if the problem does not go
away.
32010
A problem exists in the
Reseat the RS-232 board. Replace
E-RS232-ERROR-REPORTED
RS232/422 interface
the board if the problem does not go
away.
320011
E-RS232-UNAVAILABLE
A problem exists in the
RS232/422 interface
Re-seat the RS-232 board. Replace
the board if the problem does not go
away.
32009
E-CBUFFER-OVERFLOW
32012
E-NO-PENDING-MSG-BUF
General software error.
General software error.
Re-load the CCM software if this
problem recurs.
Re-load the CCM software if this
problem recurs.
Using the Patient Location
Status Messages
The following table lists messages that identify faulty conditions in the system that show up on
the CCM display. If a problem is isolated to the CCM, perform the CCM general fault isolation
procedures to further isolate the fault to the replaceable assembly.