Electrical Requirements; Line Voltage Connection; Line Voltage; Electrical Specifications - HP M2350A Service Manual

Component central monitor
Table of Contents

Advertisement

Preparing the Site
Electrical
Requirements
Electrical
Requirements
Line Voltage Connection
The HP Component Central Monitor uses a three-wire power cable, if a three-wire receptacle is
not available please consult the hospital electrician to ensure proper grounding.
Line Voltage
The HP Component Central Monitor may be operated on ac line voltage ranges of 9OV to
132V and 180V to 264V, at 5OHz or 60Hz. The voltage is selected automatically by the
switched-mode power supply. The fuse values for the system are dependent upon the line
voltage and type of display used, as detailed in the following table:
EfP Component Central Monitor Fuses
Line
Voltage
Display
Fuse
Model
Rating
100-120 v
100-120 v
200-240 V
200-240 V
14" Mono Display
14" Color Display
14" Mono Display
14" Color Display
T 2.5 A
T 4.0 A
T 2.0 A
T 3.15 A
Electrical
Specifications
Power Connections
The basic HP Component Central Monitor has only one power connector, which is located on
the Master Display module. The Master Display supplies power to the computer module which
in turn supplies the recorder modules Each slave display module added to the basic system
will have its own power connector.
Power Consumption
The power consumption of the various components of the system are as follows:
Computer Module
14" Monochrome Display
14" Color Display (M1094A)
14" Color Display (M1094B)
-
40 Watts (137 BTU)
70 Watts (239 BTU)
110 Watts (375 BTU)
86 Watts (295 BTU)
Internal
Temperature
Rise
The maximum internal temperature rise for a fully loaded computer module is less than 18°C.
5-6