HP M2350A Service Manual page 265

Component central monitor
Table of Contents

Advertisement

Checking Patient safety
Preparing
Test Equipment
'Ikble 1'1-1. Test Adapters for Conventional
Safety Checks
Type of
Equipment
A-C Adapter,
2
to 3 wire
Test Load
Test Load Parts:
Characteristics
Remove
ground pin and
connect ground terminal
with a wire to alligator or hippo clip.
Connect network between banana pins as shown in
diagram below, then connect a lead to each side of
the test load plug and install a green alligator or
hippo clip on one wire and a red one on the other
wire.
Plug
Dual banana
Resistor
1 kilohm, l%, 0.5 watt
Resistor
lo-ohm, l%, 0.5 watt
Capacitor
0.15 microfarads, 5%, 200V
Ground 'I&t Cable
Connect clip to ground wire of power cord; cut, tie
Plug
Power Cord
clip
back and insulate other wires See diagram.
Male, NEMA
6 ft (1.8 M) long, 16 gauge.
Alligator; green or black.
l0 ohms
GROUND
TEST CABLE
Figure 11-2. Test Adapters to be Made for Conventional
Safety Checks