Notes On Arrhythmia Configuration Questions - HP M2350A Service Manual

Component central monitor
Table of Contents

Advertisement

Configuring
the System
Conl?guration
Mode
Notes
on Arrhythmia
Configuration
Questions
Generate report
Record yellow
Bed Yellow Snd
Bed/f3 inop Snd
All yellow alrm
Indicates whether you want an arrhythmia report to be generated
automatically when the Arrhythmia History reaches maximum capacity.
Yes - Generates a report automatically
No - Reports must be generated manually
Indicates whether you want to record all yellow arrhythmia alarms.
Yes - Alarm is annunciated and a recording strip will run for all yellow
ZlkUTLS
No - Alarm is annunciated but no recording will be generated
Indicates whether you want the arrhythmia yellow alarm sounds to be
generated at the bedside.
Yes - Arrhythmia yellow alarms sound at the central station and at the
bedside
No - Arrhythmia yellow aIarms sound at the central station only
Indicates whether you want the
cannot
analyze inop sound to be
generated at the bedside.
Yes - inop sound at the central station and at the bedside
No - inop sound at the central station only
Indicates whether you want all arrhythmia yeIlow alarms to be
generated.
Yes - All arrhythmia yellow alarms will be generated
No - NO arrhythmia yellow alarms will be generated except Heart Rate
E-10