International Faceplates And Faceboxes - HP M2350A Service Manual

Component central monitor
Table of Contents

Advertisement

Prepasing the Site
Preparing and Terminating
Cables
Assemble faceplate and connectors to wallbox as follows:
1. Assemble faceplate to wallbox using the jackscrew with one flat washer on the front on the
faceplate and one lock washer and nut on rear of faceplate as shown in the figure below.
Caution
Tighten securely to prevent loosening of nuts
during
mating and unmating of
connectors.
2. Use screws provided with faceplate for assembling the faceplate to the wallbox.
FACEPLATE
-----
-__
LOCKWASHI
NUT
1
- -!&
-0
I
---cg.
f
NUT
/
LOCKWASHI
-------
\
WALLBOX
WALL BOX
SCREW
/
J
.-e---
---_
FLATWASHER
J
JACKSCREW
&&-
-CONNECTOR
-l-e--
\
z
JACKSCREW
FLATWASHER
-e
---_----
1
WALL BOX
SCREW
International
Faceplates and Fhceboxes
Where remote installations are required, using cables over 3m in length the terminated cables
must be assembled into faceplates and faceboxes in one of the following ways:
n
Attaching connectors to faceplates
n
Mounting faceplates into wall boxes
n
Mounting faceplates into F'aceboxes (HDLC and HP-HIL)
Note
Rack cables up to 6m long, and analog cables of 4m in length may be used
3
without connection into faceplates.