Installing
the System
Installing
the Components
5. Connect the g-pin Male connector of cable M1181A-#A23 (or #A24) to the function box.
6. Connect the AC power cord to the rear of the display.
7. Plug the power cord into the isolation transformer, or wall power outlet.
When the D2800 Display is used in the patient area or driven by video
from a function box in the patient area, power must be supplied by one of
the following
isolation
transformers:
HP M1247A Isolation
Transformer
(110 V) or HP M1964A Isolation
Transformer
(220 V).
8. If used, connect the isolation transformer power cord to the wall power outlet.
9. Switch the monitoring instrument on, then the display. See the user's guide for the display
for a display performance assurance test.