HIGH error codes and those MEDIUM and LOW error codes noted as exceptions may indicate a
software defect. Although you may not be able to permanently correct such a software defect
on-site, you may be able to temporarily clear the resulting error condition by taking corrective
action in the following order:
1. Try hot start (switch system off and on). Note that a HIGH error code will have already
generated a hot start.
2. Try cold start (change operating mode or switch system off for more than 3 hours). Note
that a cold start will unassign arrhythmia, delete configurable arrhythmia information, and
erase alarm history.
3. Reload the Arrhythmia CPC software and reprogram the EEPROM settings on the
Utility CPU Board, using factory defaults See the
HP CPC Programming
ZboZ Miinml
(M2300-90046) for more information.
4. Replace the Arrhythmia CPC in Slot 8 and program the replacement CPC.
5. Reload the Operating System software on the CPC in Slot 22.
Reporting
Arrhythmia
Error
Codes
All HIGH error codes and those MEDIUM and LQW error codes noted as exceptions should be
reported to the Hewlett-Packard Medical Response Center. Provide a complete description of
the problem, including the following information:
General Information
'
n
Frequency with which the problem occurs
n
Any user action prior to the problem's occurrence
n
What action, if any, cleared the problem
System-Generated
Information
w Contents of the Status Log
n
General Configuration of your HP Component Central Monitor
n
Arrhythmia Configuration:
q
Number of beds assigned to arrhythmia
o Number of paced patients
q
Description of patient arrhythmia
n
Arrhythmia stored alarm history for all assigned beds (if available)
n
Any related delayed/realtime recording strips
9-19