Performing Calibration With Working Standard (Only With Option 004); Measuring The Working Standard Value; Calibration With Working Standard - Agilent Technologies 4286A User Manual

Rf lcr meter
Table of Contents

Advertisement

Performing Calibration with Working Standard (only with
option 004)
Note
This procedure is necessary when you use 4286A with option 004.

Measuring the Working Standard Value

Measure the working standard to obtain L S required for LOAD compensation. Specify 51 for
resistor.
1. Prepare the test head and perform OPEN/SHORT/LOAD calibrations for the contact surface
of APC-7 R
connector.
2. Connect xture to the test head and perform OPEN/SHORT compensations. Make sure that
you performed the xture compensation with shorting device for the working standard.
3. Connect the chip resistor for working standard to the xture.
4. Measure L s of the connected chip resistor. We recommend the following conditions for
accurate and robust measurement.
Frequency
Measurement parameter
Averaging factor
OSC level
5. Record the measurements.

Calibration with Working Standard

Follow the procedure below.
1. Make sure that you complete the procedure shown above.
2. Set up calibration kit.
NNNNNNNNNNNNNNNNNNNNNNN
Press
CAL KIT ,
NNNNNNNNNNNNNNNNNNNN
Press
MODIFY .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Press
DEFINE STANDARD .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Press
LOAD RESIST. (R) .
Press
,
,
4
5
5
4
1
5
4
x1
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Press
LOAD INDUCT. (L) .
Set the value which is given by the procedure in \Measuring the Working Standard Value"
to L S .
3. Perform OPEN/LOAD/SHORT calibration procedures.
The table below lists typical values of the working standars with option 004. The values serves
as reference: use actually obtained results for calibration kit.
NNNNNNNNNNNNNNNNNNNNNNNNNN
USER KIT to toggle it
consequently to set R to 51 .
5
100 MHz
L s -R s
32
0.5 V
NNNNNNNNNNNNNN
USER .
4286A with Chip Handler
4-39

Advertisement

Table of Contents
loading