Ac Overshoot Test Method Of Implementation; Signals Of Interest - Keysight U7233A Testing Notes

Ddr1 compliance test application
Table of Contents

Advertisement

AC Overshoot Test Method of Implementation

The Overshoot test can be divided into two sub-tests: Overshoot amplitude and Overshoot area. The
purpose of this test is to verify that the overshoot value of the test signal is lower than or equal to the
conformance limit of the maximum peak amplitude allowed for overshoot as specified in the JEDEC
Standard JESD79E.
When there is an overshoot, the area is calculated based on the overshoot width. The Overshoot area
should be lower than or equal to the conformance limit of the maximum Overshoot area allowed as
specified in the JEDEC Standard JESD79E.
Figure 18

Signals of Interest

Figure 19
Based on the test definition (Write cycle only):
DDR1 Compliance Testing Methods of Implementation
AC Overshoot
AC Overshoot in Infiniium oscilloscope
Data Signal
Data Strobe Signal OR
Address Signal OR
Control Signal OR
Data Mask Control Signals
Single-Ended Signals Overshoot/Undershoot Tests
5
65

Advertisement

Table of Contents
loading

This manual is also suitable for:

U7233b

Table of Contents