Required Service Test Equipment - Keysight N5224A Service Manual

2-port and 4-port pna microwave network analyzers
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
N5224A and N5225A

Required Service Test Equipment

a
Critical Specifications
Equipment
Compression
None specified
test set
Dynamic accuracy
None specified
test set
c
N/A
Test software
Freq: 10 MHz to 20 GHz
Frequency counter
Accuracy : 0.5 ppm
Signal generator
CW Freq: 1.185 GHz
Power meter
Accuracy: ±0.0068 dB
Power sensor
Freq: 10 MHz to 4.2 GHz
Range: –30 to +20 dBm
Power sensor
Freq: 50 MHz to 40 or 50 GHz
Range: –30 to +20 dBm
Min Freq: 1 MHz
Max Freq: > 4 GHz
Spectrum analyzer
Resolution BW: 300 Hz
Digital voltmeter
Resolution: 10 mV
Printer
N/A
Mouse
N/A
Keyboard
N/A
a. Unless specified otherwise, equipment listed is required for all analyzer models.
b. P = Performance tests, A = Adjustments, T = Troubleshooting, V = System verification
c. The recommended model or part number for all equipment listed with a "P" in the Use column is required for proper operation of
this test software.
d. If an accurate measurement of the dynamic accuracy specification is not required, the E4418A or E4419A can be used.
Service Guide N5225-90001
Recommended
Model or Part
Number
Test Instruments and Software
U3070AK01
U3020AD01
N7840A
53151A
Opt 001
N5181A, 
N1913A/14A
N8482A
N8487A
E444xA PSA series,
N90xxA signal
analyzer family
Any
Any printer with Microsoft Windows XP
or Windows 7 driver
Any
Any
General Product Information
Required Service Test Equipment
Alternate Model
Use
or Part Number
None
P
None
P
None
P
None
P, A,T
E8257D, 
P, A, T
Option 520
E4418A/B,
P, A,T
d
E4419A/B
8482A
P, A,T
8487A
P, A,T
856xE
A,T
Any
T
--
Any
--
Any
--
b
2-5