Calibrating The Oscilloscope - Keysight U7233A Testing Notes

Ddr1 compliance test application
Table of Contents

Advertisement

2
Preparing to Take Measurements

Calibrating the Oscilloscope

If you haven't already calibrated the oscilloscope and probe, see
Oscilloscope and Probe.
If the ambient temperature changes more than 5 degrees Celsius from the calibration temperature,
NOTE
internal calibration should be performed again. The delta between the calibration temperature and
the present operating temperature is shown in the Utilities>Calibration menu.
If you switch cables between channels or other oscilloscopes, it is necessary to perform cable and
NOTE
probe calibration again. Keysight recommends that, once calibration is performed, you label the
cables with the channel on which they were calibrated.
26
Chapter
13, "Calibrating the Infiniium
DDR1 Compliance Testing Methods of Implementation