Single-Ended Signals Ac Output Parameters Tests; Probing For Single-Ended Signals Ac Output Parameters Tests; Test Procedure - Keysight U7233A Testing Notes

Ddr1 compliance test application
Table of Contents

Advertisement

4

Single-Ended Signals AC Output Parameters Tests

Probing for Single-Ended Signals AC Output Parameters Tests

When performing the Single-Ended Signals AC Output Parameters tests, the DDR1 Compliance Test
Application will prompt you to make the proper connections. The connection for the Single-Ended
Signals AC Output Parameters tests may look similar to the following diagram. Refer to the
Connection tab in DDR1 Compliance Test Application for the exact number of probe connections.
DDR DIMM
Figure 12
You can use any of the oscilloscope channels as Pin Under Test (PUT) source channel. You can
identify the channels used for each signal in the Configuration tab of the DDR1 Compliance Test
Application. (The channels shown in
For more information on the probe amplifiers and differential probe heads, see
Probing," starting on page 185.

Test Procedure

1
2
3
4
54
Infiniium Oscilloscope
InfiniiMax solder-in probes
Probing for Single-Ended Signals AC Output Parameters Tests with Two Probes
Start the automated test application as described in
Application"
on page 27.
Ensure that the RAM reliability test software is running on the computer system where the DDR
Device Under Test (DUT) is attached. This software will perform a test on all unused RAM on the
system by producing a repetitive burst of read-write data signals to the DDR memory.
Connect the differential solder-in probe head to the PUTs on the DDR DIMM.
Connect the oscilloscope probes to any channels of the oscilloscope.
Figure 12
are just examples).
"Starting the DDR1 Compliance Test
DDR1 Compliance Testing Methods of Implementation
Chapter
14, "InfiniiMax