Differential Signals AC Input Parameters Tests

Probing for Differential Signals AC Input Parameters Tests

When performing the Differential Signals AC Input Parameters tests, the DDR1 Compliance Test
Application will prompt you to make the proper connections. The connection for the Differential
Signals AC Input Parameters tests may look similar to the following diagram. Refer to the Connection
tab in DDR1 Compliance Test Application for the exact number of probe connections.
DDR DIMM
DDR DIMM
Figure 22
You can use any of the oscilloscope channels as Pin Under Test (PUT) source channel. You can
identify the channels used for each signal in the Configuration tab of the DDR1 Compliance Test
Application. (The channels shown in
For more information on the probe amplifiers and differential probe heads, see
Probing," starting on page 185.

Test Procedure

1
2
72
Infiniium Oscilloscope
Infiniium Oscilloscope
InfiniiMax solder-in probes
Probing for Differential Signals AC Input Parameters Tests with Three Probes
Start the automated test application as described in
Application"
on page 27.
Ensure that the RAM reliability test software is running on the computer system where the DDR
Device Under Test (DUT) is attached. This software will perform a test on all unused RAM on the
system by producing a repetitive burst of read-write data signals to the DDR memory.
I fi iiM
ld
i
b
Figure 22
are just examples).
"Starting the DDR1 Compliance Test
DDR1 Compliance Testing Methods of Implementation