To test the single-clock, single-edge, state acquisition

Testing the single-clock, single-edge, state acquisition verifies the performance of the
following specifications:
Minimum master to master clock time.
Maximum state acquisition speed.
Setup/Hold time for single-clock, single-edge, state acquisition.
Minimum clock pulse width.
This test checks the data channels using a single-edge clock at three selected
setup/hold times.
Equipment Required
Equipment
Pulse Generator
Digitizing Oscilloscope
Adapter
SMA Coax Cable (Qty 3)
BNC Cable
Coupler
BNC Test Connector,
6x2 (Qty 4)

Set up the equipment

Turn on the equipment required and the logic analyzer. Let them warm up for
1
30 minutes before beginning the test if you have not already done so.
Set up the pulse generator.
2
a Set up the pulse generator according to the following table.
Pulse Generator Setup
Channel 1
Delay: 0 ps
Width: 3.5 ns
High: -0.9 V
Low: -1.7 V
b Disable the pulse generator channel 2 COMP (with the LED off).
Critical Specifications
100 MHz 3.5 ns pulse width, < 600 ps rise time
≥ 6 GHz bandwidth, < 58 ps rise time
SMA(m)-BNC(f)
18 GHz bandwidth
BNC(m)(m) 48 in. >2 GHz bandwidth
BNC(m)(m)
Channel 2
Doub: 20.0 ns
Width: 3.5 ns
High: -0.9 V
Low: -1.7 V
Recommended Model/Part
Agilent 8131A option 020
Agilent 54121T
Agilent 1250-1200
Agilent 8120-4948
Agilent 8120-1840
Agilent 1250-0216
Period
40 ns
3–23