4

CONFigure Subsystem

CONFigure:SSI

Syntax
CONFigure:SSI <status>, (@<address_list>)
This command sets the synchronization status when used in the U2781A modular
instrument chassis.
CONFigure:SSI?
This query command returns a string value, which represents the synchronization
status.
The U2722A/U2723A is not designed to function as a master in synchronization.
NOTE
Parameters
Item
<address_list>
<status>
– This command is only operational when it is used in a U2781A modular
NOTE
– Each U2722A/U2723A occupies two slots in the U2781A.
Remark
This command will force all reference clock sources to SSI (Synchronous
Simultaneous Interface).
30
Type
Range of values
NR1
0 through 7
CPD
– NONE: None
– SLAVe: To receive triggering source from a master
instrument chassis.
Keysight U2722A/U2723A Programmer's Reference
Defaul t value
Required parameter
NONE