Replacing The Interface Board; Replacing The Detector's Firmware - Agilent Technologies G1314B VWD User Manual

Variable wavelength detector
Table of Contents

Advertisement

Replacing the Detector's Firmware

When required
Tools required
Parts required
Preparations
The installation of older firmware might be necessary:
• to keep all systems on the same (validated) revision, or
• if third part control software requires a special version.
To upgrade/downgrade the detector's firmware the following steps have to be
performed:
1 Download the module's firmware, the LAN/RS-232 FW Update Tool Version
2 Load the firmware into the detector as described in the documentation.
The G1314C VWD-SL requires firmware revision A.06.02 or above (main and resident).
N O T E
1200 Series Variable Wavelength Detector User Manual
If new version solves problems of currently installed version or after exchange
of the detector main board (VWM) the version on board is older than previous
installed one.
LAN/RS-232 Firmware Update Tool, or
Instant Pilot G4208A or Control Module G1323B
Firmware, tools and documentation from Agilent web site
Read update documentation provided with the Firmware Update Tool.
2.00 or above and the documentation from the Agilent web
http://www.chem.agilent.com/scripts/cag_firmware.asp.
Maintenance
8
99

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1200 seriesG1314c vwd-sl

Table of Contents