Adc Hardware Error; Illegal Temperature Value From Sensor On Main Board - Agilent Technologies 1200 Infinity Series User Manual

Hide thumbs Also See for 1200 Infinity Series:
Table of Contents

Advertisement

7
Error Information
Detector Error Messages

ADC Hardware Error

Error ID: 7830, 7831
A/D- Converter hardware is defective.
Probable cause
1

Illegal Temperature Value from Sensor on Main Board

Error ID: 1071
This temperature sensor (located on the detector main board) delivered a
value outside the allowed range. The parameter of this event equals the
measured temperature in 1/100 centigrade. As a result the temperature
control is switched off.
Probable cause
1
2
110
A/D-Converter hardware is defective.
Defective sensor or main board.
Detector is exposed to illegal ambient
conditions.
Suggested actions
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Verify that the ambient conditions are within
the allowed range.
Agilent 1200 Infinity Series VWD - User Manual