Replacing Capillaries On A Standard Flow Cell - Agilent Technologies 1100 Series User Manual

Diode array and multiple wavelength detector sl
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Replacing Capillaries on a Standard Flow Cell

When required
Tools required
Parts required
Preparations
1
Identify the inlet and outlet capillaries. To replace the inlet
capillary, continue with step 3.
1100 Series DAD and MWD User Manual
If the capillary is blocked
Two 1/4 inch wrenches for capillary connections
Wrench 4 mm for capillary connections
Screwdriver Pozi 1 PT3
For parts see
"Standard Flow Cell"
Turn the lamp(s) off.
Remove the front cover.
Remove the flow cell, see
Notes
The fittings at the flow cell body are special types for
low dead volumes and not compatible with other
fittings.
When retightening the fittings, make sure that they
are carefully tightened (handtight plus 1/4 turn with
a wrench). Otherwise damage of the flow cell body
or blockage may result.
Inlet
Outlet
on page 156
"Exchanging a Flow Cell"
Maintenance
8
on page 121.
131