7
Error Information
General Error Messages
Leak Sensor Short
Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current increases above the upper limit, the
error message is generated.
Probable cause
1
2
3
4
Leak Sensor Open
Error ID: 0083
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current falls outside the lower limit, the error
message is generated.
Probable cause
1
2
3
4
Agilent InfinityLab LC Series Diode Array Detectors User Manual
Defective leak sensor.
Leak sensor incorrectly routed, being
pinched by a metal component.
Power switch assembly defective
Cable or contact problem.
Leak sensor not connected to the power
switch board.
Defective leak sensor.
Leak sensor incorrectly routed, being
pinched by a metal component.
Power switch assembly defective
Suggested actions
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Suggested actions
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
112