Section Two: Calibrated Leak Replacement for VS Series Leak Detectors

Equipment Required
Extended Length M5 Allen Wrench (included with leak detector)
Slotted Screw Driver
Installation Procedure
For clarity, some items have been omitted from views.
WARNING
1. Turn off the power switch located on the back of the unit and unplug.
2. Wait 30 seconds for the high voltage to dissipate.
3. Using an extended length M5 Allen wrench, remove the four screws holding the rear plastic cover (Figure 1:
Rear Cover Screws) and detach the rear plastic cover from the unit.
Calibrated Leak Replacement for VS C15, VS, and (G8610, G8611, or G8612) Series Leak Detectors
Figure 1: Rear Cover Screws
Disconnect power from the unit before performing any maintenance procedure that
requires physically disconnecting any part of the system.