Exchanging the Leak Sensor

When required
Tools required
Parts required
Do not switch on the detector when the cover is removed.
WA R N I N G
Electronic boards and components are sensitive to electronic discharge (ESD). In order
C A U T I O N
to prevent damage always use an ESD protection (for example, the ESD wrist strap
from the accessory kit) when handling electronic boards and components, see
the ESD Strap"
1 Turn off the lamp.
2 Switch off the detector and disconnect the cables.
3 Remove detector from stack and place it on the working bench.
4 Remove the front cover, top cover and top foam section, see
5 Remove the optical unit, see
6 Disconnect the leak sensor cable from the detector main board (J21).
On the G1315B DAD and G1365B MWD the leak sensor can be pulled out without the
N O T E
removing the leak pan (the next steps describe the procedure for the G1315A/G1365A).
1100 Series DAD and MWD Reference Manual
If defective
Screwdriver POZI 1 PT3
Leak sensor assembly 5061-3356
on page 98.
Top Cover and Foam"
(G1315A/G1365A)"
(G1315B/G1365B)"
on page 156.
on page 140.
"Removing the Optical Unit
on page 154 or
"Removing the Optical Unit
Repairing the Detector
"Using
"Removing the
4
167