Lost CAN Partner

Error ID: 0071
During an analysis, the internal synchronization or communication between
one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one or
more of the modules is no longer recognized as being connected to the system,
the error message is generated.
Probable cause
1
2
3

Leak Sensor Short

Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current increases above the upper limit,
the error message is generated.
Probable cause
1
2
Agilent 1200 Infinity Series DAD User Manual
CAN cable disconnected.
Defective CAN cable.
Defective main board in another module.
Defective leak sensor.
Leak sensor incorrectly routed, being
pinched by a metal component.
Error Information
General Error Messages
Suggested actions
• Ensure all the CAN cables are connected
correctly.
• Ensure all CAN cables are installed
correctly.
Exchange the CAN cable.
Switch off the system. Restart the system, and
determine which module or modules are not
recognized by the system.
Suggested actions
Please contact your Agilent service
representative.
• Please contact your Agilent service
representative.
• Correct the routing of the cable.
• If cable defective, exchange the leak sensor.