Transducer Surface Temperature Limits

The following table provides the maximum surface temperature of the
transducers compatible with the system.
Maximum surface temperatures are in accordance with IEC 60601-2-37.
Transducer
C5-2
C6-2
C6-3 3D/C6F3
C8-5
5.0C50+
MPT7-4
BE9-4
EV9-4
EC9-4
Endo-V II
Endo-V 3D/EV8F5
5.0L45
7.5L70
L10-5
VF13-5
VF13-5SP
P4-2
5.0P10
CW2
CW5
TMM = Tissue Mimicking Material
[ 1 ]
I N S T R U C T I O N S
Maximum Temperature
TMM
41.0 C
41.0 C
41.0 C
41.0 C
41.0 C
41.0 C
41.4 C
41.0 C
41.0 C
41.0 C
41.0 C
41.0 C
41.0 C
41.0 C
41.1 C
41.8 C
41.0 C
41.1 C
41.0 C
41.0 C
F O R
U S E
2 S a f e t y a n d C a r e
[2] Instructions for Use
MPT7-4
Temperature
Controls
Still Air
41.7 C
41.4 C
41.0 C
41.0 C
41.3 C
41.0 C
41.9 C
41.0 C
41.0 C
41.0 C
41.0 C
41.0 C
41.0 C
41.0 C
41.0 C
42.9 C
41.0 C
41.6 C
41.0 C
41.0 C
Ch C3
2 - 1 3