6 - 2

Power board D2

WARNING
Disconnect boards D1 and D2 completely.
Disconnect the cables between boards D1 and D2.
Remove board D1 and D2.
Apply heat conducting paste to the heat sink for the power semiconductor on the new D2
board.
Reinstall board D2 and D1 and reattach all connections.
Ensure that the shielding and ground connections are positioned correctly.
Perform the kV offset adjustment and the generator adjustment (filament learning).
Test the FL / PFL / DR and direct exposure functions, if there is a cassette holder .

Interface board D30

Replace interface board D30.
Check the +27 V /+24 V /+24 V voltage. Adjust it , if necessary. Refer to chapter 3 of
these instructions, Checking the operating voltages.
Check the radiation release / format switch over / vertical column travel / brakes func-
tions.
SIREMOBIL Iso-C
System Manual
Replacing boards / replacing components
Electrical voltage!
See Chapter 1, Safety Information.
Make sure that LED V400 is off prior to performing any work on
boards D1 and D2. This should occur approximately 3 minutes
after switching the SIREMOBIL Compact OFF.
SPR2-230.061.01
Rev. 13
Page 2 of 48
12.04