Checking / Replacing the Lifting Columns R / L
Removing the lifting column
•
Disconnect the plug from the actual value acquisition of the lifting columns (board D31
X10/ X11).
•
Disconnect the motor power supply plug.
•
Unlock and remove the motor encoder plug.
Fig. 24: Allen_screws_A
•
Turn the lifting column drive shaft by hand until the Allen screws of the drive shaft are
accessible from above
•
Remove the 2.5 mm Allen set screws
If necessary, remove "Loctite" using a heat gun.
•
Remove the mounting screws on the base of the lifting column (Fig. 4) and remove the
column.
NOTE
•
Remove the drive shaft bearing to replace the cables
•
Remove the strain relief on the two cables for the lifting column.
•
Remove the mounting screws on the base of the lifting column
move the column.
Siemens AG
Medical Solutions
((Fig. 24 / p.
Ensure that the encoder on the motor shaft does not become dam-
aged.
RXL2-120.841.02.01.02
09.05
Fig. 25: Allen_screws_B
59)).
((1/Fig. 24 / p. 59)
and
((Fig. 26 / p.
Page 59 of 72
CS PS 24
(1/Fig. 25 / p.
59)).
60)).
((Fig. 24 / p.
59)) and re-
LITHOSTAR Multiline
59