7.4
Setting the RJC Function on Channels Set to
TC Input
Procedure
7-6
This section explains the details of setting RJC on channels set to TC input.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Changing the Settings
1.
Hold down the
MENU
2.
Hold down the
Setting mode.
3.
Press the
key or SHIFT +
4.
Set each item and press the
Use the
key or SHIFT +
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up to
then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Basic=RJC
CH=01-01
First channel Last channel
RJC=Internal
(When RJC is set to External)
Volt=
01-01 Channel
ESC/?
Setting complete
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
the
key to select Store and then press the
the screen returns to Operation mode. If you select Abort and press the
changes are discarded, and the screen returns to Operation mode. Press the ESC key to
return to the Basic= screen.
key for 3 seconds to enter Setting mode.
and
keys simultaneously for 3 seconds to enter Basic
key to select RJC and then press the
key.
key to select values.
Set the channel range.
Select the RJC mode of the
thermocouple input.
Set the compensation voltage when
0
using the external RJC function.
The new setting takes effect.
key to select End and then press the
key. The changes are applied, and
key.
key. Press
key, the
IM 04P01B01-01E