7.15 Enabling the Bias, Low-Cut, Alarm Delay, Calibration Correction (/CC1 Option) Functions
Explanation
7-24
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
the
key to select Store and then press the
the screen returns to Operation mode. If you select Abort and press the
changes are discarded, and the screen returns to Operation mode. Press the ESC key to
return to the Basic= screen.
Bias
Enables/Disables the bias function.
Use:
Enables the setting of the bias value in Setting mode.
Not:
The Bias item does not appear in Setting mode.
SQRT low-cut
Enables/Disables the square root low-cut function.
Use:
Enables the setting of the low-cut function when a channel is set to square root
computation in Setting mode.
Not:
The SQRT low-cut item does not appear in Setting mode.
1-5V low-cut
Enables/Disables the 1-5V low-cut function.
Use:
Enables the setting of the low-cut function when a channel is set to 1-5V in
Setting mode.
Not:
The 1-5V low-cut item does not appear in Setting mode.
Alarm delay
Enables/Disables the alarm delay function.
Use:
Enables the setting of delay high limit alarm or delay low limit alarm for the alarm
type.
Not:
The delay high limit alarm and delay low limit alarm items do not appear as
possible selections for the alarm type.
Calibration correction
Enables/Disables the calibration correction function.
Use:
Enables the setting of the correction mode and the number of calibration points
in Basic Setting mode and the setting of the correction points and correction
values in Setting mode.
Not:
The calibration item does not appear in Basic Setting or Setting mode.
Note
The bias and calibration correction functions (/CC1 option) cannot be enabled simultaneously.
<Related Topics> Setting the bias value: Section 6.12
Setting the low-cut value to the measured value: Section 5.1
Setting the delay high/low limit alarm: Section 5.2
Setting the correction mode and the number of calibration points for
calibration correction: Section 7.23
Performing calibration correction: section 6.13
key to select End and then press the
key. The changes are applied, and
key. Press
key, the
IM 04P01B01-01E