YOKOGAWA MuR10000 Manual page 37

Table of Contents

Advertisement

IM 04P01B01-01E
• Turning ON/OFF the Periodic Printout
Periodic printout can be turn ON/OFF. By default, periodic printout is enabled with the
interval synchronized to the chart speed. Also, the measured values that are printed
are the instantaneous values.
Manual Printout
The current measured values and alarm status as well as computed results when
computation mode is on can be printed with keys.
When manual printout is executed, trend recording stops and restarts when manual
printout is complete.
Message Printout
Printout is not performed when the chart speed is as follows:
Pen model: 1600 mm/h or higher, dot model: 101 mm/h or higher
Preset messages can be printed on the chart paper using the keys. Five messages, each
within 16 characters, can be registered in advance.
• If message printout is executed while another message is being printed, the most
recent message is temporarily stored to the buffer memory in a printout-wait condition.
Messages are cleared from the buffer memory when they are printed.
• The number of messages that can be stored in the buffer is 5. If message printout is
executed when the buffer is full, the message is not printed. A buffer overflow mark is
printed when there are messages that cannot be printed because the buffer is full.
• The time printout format can be selected.
New Chart Speed Printout
Printout is not performed when the chart speed is as follows:
Pen model: 1600 mm/h or higher, dot model: 101 mm/h or higher
• When the chart speed is changed, the time tick (dot model), the date/time of change,
and the new chart speed are printed. An asterisk (*) shows there are messages that
cannot be printed.
• The time printout format can be selected.
Recording Start Printout
Printout is not performed when the chart speed is as follows:
Pen model: 1600 mm/h or higher, dot model: 101 mm/h or higher
When recording is started, the time tick (dot model), the time, and the chart speed can be
printed. An asterisk (*) shows there are messages that cannot be printed.
• The recording start printout can be enabled or disabled. By default, the recording start
printout is disabled.
• The time printout format can be selected.
Printout/Display Format of the Date (Version 1.11 or Later)
The printout/display format of the date can be selected from the list below. The setting
applies to all dates printed and shown on the display. However, this does not apply to
the format of the date printout of the message printout that include the measured values
(/BT1 option).
Type
Printout/Display Format
Y/M/D
2005/08/31
M/D/Y
08/31/2005
D/M/Y
31/08/2005
D.M.Y
31.08.2005
M.D.Y
Aug.31.2005
<Related Topics> Setting the printout/display format of the date: Section 7.19
For the procedure to set the functions, see section 1.10, "Function Setup Guide."