Square Root Computation
IM 04P01B01-01E
Note
•
The displayable/printable range of scale values is –19999 to 30000 excluding the decimal.
•
If the leftmost value of the scale is set to –20000 (excluding the decimal), the value when
the low-cut function is applied is displayed as "–Over."
Set=Range
CH=01-01
First channel
Last channel
Mode=SQRT
Range=2V
Span_L= -2.000
Span_R=
2.000
Scale_L=
Scale_R= 200.00
(When low-cut is set to Use)
Low-cut=Off
(When low-cut is On)
Low-cut point=0.0
01-01 Channel
ESC/?
Setting complete
Description
• Input Type in Which Square Root Computation Can Be Performed
Square root computation can be performed when set to DC voltage input.
• Span Left and Span Right
Same as the selectable span range of the DC voltage input.
Make sure that span left is less than span right.
Set the channel range.
Select SQRT.
Select the range type of DC voltage.
20 mV, 60 mV, 200 mV, 2 V, 6 V, 20 V,
or 50 V
Set the left span value.
Set the right span value.
Set the decimal place
0.00
and left scale value.
Set the right scale value.
Turn On or Off the low-cut function.
Set the low-cut point.
0.0 to 5.0% of the recording span, 0.1 steps.
The new setting takes effect.
5.1 Setting the Input Range