Power Supply Wiring
Precautions to Be Taken While Wiring the Power Supply
Item
Rated supply voltage
Allowable power supply voltage range
Rated power supply frequency
Allowable power supply frequency range 50/60 Hz ± 2%
Maximum power consumption
IM 04P01B01-01E
Make sure to follow the warnings below when wiring the power supply. To prevent electric
shock and damage to the recorder, observe the following warnings.
Panel Mount Type
• To prevent electric shock when wiring, ensure the main power supply is turned
OFF.
• To prevent the possibility of fire, use 600 V PVC insulated wire (AWG 20 to 16)
or an equivalent wire for power wiring.
• Make sure to earth ground the protective earth terminal through minimum
resistance before you turn on the power.
• Use crimp-on lugs (designed for 4 mm screws) for power and ground wiring
termination.
• To prevent electric shock, make sure to close the transparent cover for the
power supply wires.
• Make sure to provide a power switch (double-pole type) on the power supply line
in order to separate the recorder from the main power supply. Put an indication
on this switch as the breaker on the power supply line for the recorder and
indications of ON and OFF.
Switch specifications
Rated power current: 1 A or more (other than /P1), 3 A or more (/P1 option)
Rated rush current: 60 A or more (other than /P1), 70 A or more (/P1 option)
Complies with IEC 60947-1, 3.
• Connect a fuse in the power supply line. Use a fuse approved by CSA (for the
use in North America) or VDE (for the use in Europe).
2 A to 15 A (other than /P1), 4 A to 15 A (/P1 option)
• Do not add a switch or fuse to the ground line.
Portable Type (/H5x Option)
• Ensure that the source voltage matches the rated power supply voltage of the
instrument before connecting the power cord.
• Connect the power cord after checking that the power switch of the portable type
is turned OFF.
• To prevent electric shock or fire, be sure to use the power cord for the portable
type supplied by YOKOGAWA.
• Make sure to connect protective earth grounding to prevent electric shock.
Connect the power cord of the portable type to a three-prong power outlet
equipped with a protective earth terminal.
• Do not use an extension cord that does not have a protective grounding wire.
The protective features of the instrument will be rendered ineffective.
Use a power supply that meets the following conditions:
Note
Do not use a supply voltage in the range 132 to 180 VAC, as this may have adverse effects on
the measurement accuracy.
WARNING
Power Supply Specifications
(Options Other Than /P1)
100 to 240 VAC
90 to 264 VAC
50/60 Hz
40 VA
(/P1 Option)
24V DC/AC
21.6 V to 26.4 VDC/AC
50/60 Hz (for AC)
50/60 Hz ± 2% (for AC)
25 VA (for DC) or 35 VA (for AC)