YOKOGAWA MuR10000 Manual page 146

Table of Contents

Advertisement

7.1 Changing the Auxiliary Alarm Function
Explanation
7-2
Basic=Alarm
Diagnosis=Off
Reflash=Off
AND=None
Act=Energize
Behavior=Nonhold
Indicator=Nonhold
Increase=01
Decrease=01
Hysteresis=0.5%
(models with /M1)
M_Hysteresis=0.0%
Alarm
ESC/?
Setting complete
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
the
key to select Store and then press the
the screen returns to Operation mode. If you select Abort and press the
changes are discarded, and the screen returns to Operation mode. Press the ESC key to
return to the Basic= screen.
Diagnosis
When set to On, alarm output relay I01 becomes a dedicated diagnosis output relay, and
the operation is fixed to OR, de-energized, and non-hold. Diagnosis output is a function
which outputs a relay signal when there is an error in the plotter operation on the pen
model, when a burnout is detected, or when there is an error in the A/D converter.
Select On to set relay I01 to diagnosis
output.
Select On to set relays I01, I02, and I03
to reflash alarm operation.
Select the range of relays that are to
operate using AND logic.
Select the relay's energized or
de-energized operation.
Select the relay's hold or non-hold
operation.
Select the relay's hold or non-hold
operation.
Set the interval of the high limit on
rate-of-change alarm.
Set the interval of the low limit on
rate-of-change alarm.
Set the hysteresis of the measurement
channels.
Set the hysteresis of the computation
channels.
The new setting takes effect.
key to select End and then press the
key. The changes are applied, and
key. Press
key, the
IM 04P01B01-01E