YOKOGAWA MuR10000 Manual page 22

Table of Contents

Advertisement

1.2 Measuring Input Section
Burnout Detection of Thermocouples
Reference Junction Compensation of Thermocouple Input
1-4
• Bias
A given value (bias value) is added to the input value and used as the measured value
of that channel.
Biased channel
Input value
Bias value
<Related Topics> Setting the bias: Sections 7.15 and 6.12
• Calibration Correction (/CC1 Option)
Corrects the measured value of each channel using segment linearizer approximation
and makes the resultant value the measured value of the channel. You can set
arbitrary correction values for 2 to 16 points of arbitrary measured values. Linear
approximation is used between two segment points. Correction values can be
assigned using revise values or absolute values.
Scale value
B5
Measured value
A4
Correction
value
B3
B4
B2
A3
B1
A2
A1
<Related Topics> Setting the calibration correction function: Sections 7.15, 7.23, and
6.13
This function makes the recording go off the scale to the right or left when the
thermocouple burns out while measuring temperature with a thermocouple. This function
can also be used on 1-5V. The burnout detection function can be set for each channel.
By default, this function is disabled.
Note
For 1-5V, a burnout occurs when the input value is less than or equal to 0.2 V.
<Related Topics> Setting the burnout detection function: Section 7.3
When measuring the temperature using a thermocouple, the reference junction
compensation on the recorder can be used. When using external reference junction
compensation, you can set the reference voltage. The reference junction compensation
can be set for each channel.
By default, the recorder is configured to use the internal reference junction compensation
function.
For the procedure to set the functions, see section 1.10, "Function Setup Guide."
+
Correction using
A5
revise values
Measured value = A
Correction value = B – A
A1 to A5: Measured value (measured value before correction)
B1 to B5: Correction value (measured value after correction)
Calibration point
Measured value
Correction using
absolute values
Measured value = A
Correction value = B
IM 04P01B01-01E

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents