10.2 Troubleshooting Flow Charts

Nothing works (pen and chart paper do not move).
• The reading error is large.
• The indication is unstable.
• The pen is off the scale on either the 0% or 100% side.
10-4
Follow the flow charts to correct the problem.
Is the power switch
turned ON? Does the switch
work properly?
YES
Is the power supply
connected properly?
YES
Is the power supplied
properly?
YES
Contact your nearest
YOKOGAWA dealer.
Does the input meet
the specifications?
YES
Are the span and range
settings correct?
YES
To the next page
NO
Turn ON the power properly.
NO
Connect properly.
Supply the power at the specified
NO
voltage and frequency.
NO
Change the input to meet
the specifications.
NO
Adjust the span and range.
Section 2.6
Section 2.5
IM 04P01B01-01E