12.5 Specifications of Optional Functions
Item
Measurement and recording accuracy: See "Standard Performance" on page 12-20.
Thermocouple
(excludes RJC
accuracy)
RTD
Reference junction
compensation accuracy
Remote Control 5 Points (/R1)
Item
Number of input terminals 5
Input type
Input type and signal level
Voltage-free contact
Open collector
Allowable input voltage
Signal type
Operation
Portable Type (/H5x)
Item
Installation
External dimensions
Weight
Standard accessories*
12-16
Specifications
Input Type
0 to 450°C
450 to 750°C
PR40-20
750 to 1100°C
1100 to 1900°C
PLATINEL
NiNiMo
0 to 400°C
W/WRe26
400 to 2400°C
Type N (AWG14)
0 to 20 K
Kp vs Au7Fe
20 to 300 K
Pt25
Pt50
Ni100 (SAMA)
Ni100 (DIN)
Ni120
0 to 40 K
J263*B
40 to 300 K
Cu53
Cu100
When input terminal temperature balanced (60 minutes after power on)
PLATINEL, NiNiMo, W/WRe26, or Type N (AWG14): ±1.0°C
No RJC for PR40-20 and Kp vs Au7Fe (fixed to 0°C)
Specifications
Isolated from the main circuitry through a photocoupler, built-in isolated power supply for the input
terminals, and shared common
Contact closed at 200 Ω or less and contact open at 100 kΩ or greater
ON voltage: 0.5 V or less (sink current 30 mA or more), leakage current when OFF: 0.25 mA or less
5 VDC
Edge, level, or trigger (250 ms or more)
Perform the following control operations by applying a given signal to the remote signal input terminal.
• Start/stop recording (edge)
• Alarm ACK (trigger)
• Set the clock (adjust the clock to the nearest hour with the contact input, trigger)
Time When Signal Is Input Procedure
00 min 00 s to 01 min 59 s
58 min 00 s to 59 min 59 s
02 min 00 s to 57 min to 59 s No operation
• Manual printout (trigger)
• Message 1 to 5 printout (trigger)
• Switch the chart speed (level)
• Start/stop computation (/M1 option) (edge)
Specifications
Desktop type (with a carrying handle)
144 (W) × 187 (H) × 285 (D) mm (including the handle and feet)
Add 1.0 kg to the weight given in General Specifications (see section 12.6, excluding the power cord)
With a power cord
* Panel mounting brackets are not included.
Measurement Accuracy
Not warranted
±(0.9% of rdg + 3.2°C)
±(0.9% of rdg + 1.3°C)
±(0.9% of rdg + 0.4°C)
±(0.25% of rdg + 2.3°C)
±(0.25% of rdg + 0.7°C)
±15.0°C or less
±(0.2% of rdg + 2.0°C)
±(0.2% of rdg + 1.3°C)
±4.5 K
±2.5 K
±(0.15% of rdg + 0.6°C)
±(0.3% of rdg + 0.6°C)
±(0.15% of rdg + 0.4°C)
±3.0 K
±1.0 K
±(0.15% of rdg + 0.8°C)
±(0.2% of rdg + 1.0°C)
Truncates the minutes and seconds.Example: 10:00:50 → 10:00:00
Rounds up the minutes and seconds.Example: 10:59:50 → 11:00:00
• Reset computation (/M1 option) (trigger)
• Priority to remote recording (/BT1 option) (edge)
• Batch comment switching (/BT1 option) (level)
• Start/stop data saving (/EM1 option) (edge)
Recording Accuracy
Measurement accuracy
±(0.3% of recording span)