6.12 Applying a Bias on the Measuring Input Signal

Procedure
Explanation
IM 04P01B01-01E
This section explains the details of setting a bias on the measuring input signal.
If you change the input range, set the bias again.
1.
Hold down the
key for 3 seconds to enter Setting mode.
MENU
2.
Press the
key or SHIFT +
3.
Set each item and press the
Use the
key or SHIFT +
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up to
then are cancelled, and the display returns to a higher level menu.
4.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
5.
Hold down the
key for 3 seconds to return to Operation mode.
MENU
Set=Bias
CH=01-01
First channel
Last channel
Bias=Off
(When Bias is On)
Bias=
0.000
01-01 Channel
ESC/?
Setting complete
Channel Range
The only case when a bias can be set on multiple channels simultaneously is when the
channels are set to the same range type (for example channel 01 and 02 are set to 2 V
range). For channels on which scaling is set, the channels must be set to the same range
type, same span values, and same scaling values.
Bias
The range of bias that can be specified is ±10% of the span of the measurable range at
the specified range type. For example, the range is –0.4 V to 0.4 V for the range type 2 V.
For channels on which scaling is set, the range is ±10% of the scaling span.
Note
A bias cannot be set on channels set to ON/OFF input (DI).
<Related Topics> Enabling the bias function: Section 7.15
key to select Bias and then press the
key.
key to select values.
Set the channel range.
Select whether to use the bias.
Set the bias value.
The new setting takes effect.
1
2
3
key.
4
5
6
7
8
9
10
11
12
App
Index
6-13