9.6 Specifying the Timer Used in Statistical Calculations (TLOG)
Explanation
9-14
Timer No.
There are three timers.
Periodic:
The timer for determining the periodic printout interval. The computed result
is printed every interval, and the computed TLOG value is reset.
1:
The timer of timer No. 1. Specify whether the computed value is printed every
interval and whether to reset the computed value.
2:
The timer of timer No. 2. Specify whether the computed value is printed every
interval and whether to reset the computed value.
• Timer operation
The table below shows how each timer operates.
Item
Application
Target channel
Timer mode**
Reset computed TLOG value
Print when computation stops
* For the timer 1 and 2 settings, see section 9.12.
** Absolute time mode:
Relative time mode:
SUM scale
Set the sum scale when using statistical calculation (TLOG.SUM).
For the definition of the term "sum scale," see page 7-14.
Off:
Simply sums the computed values.
/s:
Sums by converting the computed values to a value over 1 second.
/min:
Sums by converting the computed values to a value over 1 minute.
/h:
Sums by converting the computed values to a value over 1 hour.
/day:
Sums by converting the computed values to a value over 1 day.
<Related Topics> Setting the periodic printout timer: Section 7.8
Setting Timer 1 and Timer 2: Section 9.12
Timer #
Periodic
Periodic
Measurement and
computation channels
Absolute time mode
Yes
No
Keeps time from the reference time.
Starts the time in sync with the starting of the computation.
1 and 2*
TLOG computation
Computation channels
Absolute or relative time mode
Yes/No selectable
Yes
IM 04P01B01-01E