YOKOGAWA MuR10000 Manual page 109

Table of Contents

Advertisement

Linear Scaling
IM 04P01B01-01E
Set=Range
CH=01-01
First channel
Last channel
Mode=Scale
Type=Volt
Range=2V
Span_L= -2.000
Span_R=
2.000
Scale_L=
Scale_R= 200.00
01-01 Channel
ESC/?
Setting complete
Description
• Scale Left, Scale Right, and Decimal Place
Selectable range (mantissa): –20000 to 30000
Example: T he value in the range of –100.00 to 350.00 cannot be specified. The
mantissa of Scale_R is 35000, which exceeds the upper limit of 30000. Set
in the range of –100.0 to 350.0.
Decimal place: XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX
The decimal place is set using Scale_L. It cannot be set using Scale_R.
Example: Change 0.00 to 0.0.
Press the
Press the
Note
The displayable/printable range of scale values is –19999 to 30000 excluding the decimal.
<Related Topics> Setting the unit: Section 5.3
Set the channel range.
Select Scale.
Select Volt, TC, RTD, or DI.
Select the range type.
Set the left span value.
Set the right span value.
Set the decimal place
0.00
and left scale value.
Set the right scale value.
The new setting takes effect.
key to move the cursor to the last digit of 0.00.
key to change the value to 0.0 .
Space
5.1 Setting the Input Range