Chapter 12
Specifications
12.1 Input Specifications
Number of Inputs and Scan Interval on the Pen Model
Item
Number of inputs
Scan interval
Number of Inputs and Scan Interval on the Dot Model
Item
Number of inputs
Scan interval
Input Type
Item
Input type
Input Type and Measurable Range
IM 04P01B01-01E
Specifications
1, 2, 3, or 4
125 ms
Specifications
6
1 s (when the integration time of the A/D converter is 20 ms or 16.7 ms)
2.5 s (when the integration time of the A/D converter is 100 ms)
Specifications
DC voltage, 1-5V, TC, RTD, DI (ON/OFF input), DC current (by adding an external shunt resistor)
Input
Range Type
Volt
20 mV
60 mV
200 mV
2 V
6 V
20 V
50 V
1-5V
1
TC
R
S
1
1
B
1
K
1
E
1
J
1
T
1
N
W
2
3
L
3
U
4
WRe
5
RTD
PT (Pt100)
5
JPT (JPt100)
DI
Level
Contact
6
1
R, S, B, K, E, J, T, N: IEC 584-1 (1995), DIN IEC 584, JIS C1602-1995
2
W: W-5% Re/W-26% Re (Hoskins Mfg. Co.), ASTM E988
3
L: Fe-CuNi, DIN 43710, U: Cu-CuNi, DIN 43710
4
WRe: W-3%Re/W-25%Re (Hoskins Mfg. Co.)
5
Pt100: JIS C 1604:2013, IEC 60751:2008, DIN EN 60751:2009
JPt100: JIS C1604-1989, JIS C1606-1989
Measurement current: i = 1 mA (Pt100, JPt100)
6
Detected current value: approx. 10 μA
Measurable Range
–20.00 to 20.00 mV
–60.00 to 60.00 mV
–200.0 to 200.0 mV
–2.000 to 2.000 V
–6.000 to 6.000 V
–20.00 to 20.00 V
–50.00 to 50.00 V
(0.800 to 1.200 V range) to (4.800 to 5.200 V range)
0.0 to 1760.0°C
0.0 to 1760.0°C
0.0 to 1820.0°C
–200.0 to 1370.0°C
–200.0 to 800.0°C
–200.0 to 1100.0°C
–200.0 to 400.0°C
0.0 to 1300.0°C
0.0 to 2315.0°C
–200.0 to 900.0°C
–200.0 to 400.0°C
0.0 to 2400.0°C
–200.0 to 600.0°C
–200.0 to 550.0°C
0: Less than 2.4 V, 1: Greater than or equal to 2.4 V
Within ±6 V
0: Open, 1: Closed
32 to 3200°F
32 to 3200°F
32 to 3308°F
–328 to 2498°F
–328.0 to 1472.0°F
–328.0 to 2012.0°F
–328.0 to 752.0°F
32 to 2372°F
32 to 4199°F
–328.0 to 1652.0°F
–328.0 to 752.0°F
32 to 4352°F
–328.0 to 1112.0°F
–328.0 to 1022.0°F
12-1
1
2
3
4
5
6
7
8
9
10
11
12
App
Index