7.2
Changing the Integration Time of the A/D
Converter
Procedure
Explanation
7-4
This section explains the details of setting the integration time of the A/D converter.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Changing the Settings
1.
Hold down the
MENU
2.
Hold down the
Setting mode.
3.
Press the
key or SHIFT +
4.
Set each item and press the
Use the
key or SHIFT +
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up to
then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
If you are done, press the ESC key.
Basic=A/D
Integrate=Auto
A/D Integrate
ESC/?
Setting complete
Applying the Changes and Returning to Operation Mode
key or SHIFT +
Press the
the
key to select Store and then press the
the screen returns to Operation mode. If you select Abort and press the
changes are discarded, and the screen returns to Operation mode. Press the ESC key to
return to the Basic= screen.
Integrate
Selects the integration time of the A/D converter.
Setting
Integration Time
50 Hz
20 ms
60 Hz
16.7 ms
Auto
20 ms or 16.7ms
1
100 ms
100 ms
1: Selectable only on the dot model.
2: Automatically switch by detecting the power supply frequency.
3: The integration time is fixed to 20 ms if "Auto" is selected when using the DC power supply on a
model with the /P1 option.
key for 3 seconds to enter Setting mode.
and
keys simultaneously for 3 seconds to enter Basic
key to select A/D and then press the
key.
key to select values.
key.
Select the integration time of the
A/D converter.
The new setting takes effect.
key to select End and then press the
Scan Interval (Pen Model)
125 ms
125 ms
2, 3
125 ms
key. The changes are applied, and
Scan Interval (Dot Model)
1 s
1 s
1 s
2.5 s
key.