9.5

Setting the Alarm

Procedure
Set=Math
IM 04P01B01-01E
This section explains the details of setting alarms for each computation channel.
If you change the computing equation or the recording span, set the alarm again.
1.
Hold down the
key for 3 seconds to enter Setting mode.
MENU
2.
key or SHIFT +
Press the
3.
Press the
key or SHIFT +
key.
4.
Set each item and press the
Use the
key or SHIFT +
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up to
then are cancelled, and the display returns to a higher level menu.
5.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
6.
Hold down the
key for 3 seconds to return to Operation mode.
MENU
Math=Alarm
CH=0A-0A
First channel
Last channel
Level=1
Alarm=Off
Type=H
Value=
Relay=Off
Relay No.=I01
0A-0A CH/level 1
ESC/?
Setting complete
key to select Math and then press the
key to select Alarm and then press the
key.
key to select values.
Set the channel range.
Select the level number between 1 and 4.
Select On to set the alarm. When set
to On, the settings below can be entered.
Select the alarm type.
Select the value used to turn On the alarm.
0.00
Turn relay On when outputting alarms on
the relay. This setting is valid only on
models with the /A1, /A2, or /A3 option.
Specify the alarm output relay No.
The new setting takes effect.