Replacing the Fuse
Agilent N4373A Lightwave Component Analyzer, First Edition
1 On the rear of the optical test set, use a flat- tipped
screwdriver to lever open the fuseholder
2 Pull the fuse holder and fuse clear of the back panel.
3 Take the fuse out of the fuse cap and note the fuse rating.
4 Make sure the replacement fuse has the same rating.
5 Put the fuse into the fuse holder.
6 Put the fuse and fuse cap into the receptacle in the rear
of the case and press it into place.
Troubleshooting and Maintenance
(1)
(1).
93