Source Pretune Correction Constants (l&t 43)
Required Equipment and 'Ibols (HP 8719D/20D only)
Frequency counter . . . . . . . , . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 535OB'
RF cable . . . . . . . . , . . . . . . . . . . . . . . . . . . . , . . . . . . . . . . . . . . . . . . . . . . . . . .HP 85131C/D/E/F
*The frequency accuracy of the HP 8566B/633 spectrum analyzer is suflkient
for this procedure.
Frequency counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . , . . . . . . . . . . . . . HP 5351B*
RF cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 85133(2/D/E/F
*The frequency accuracy of the HP 8566B/633 spectrum analyzer is sufficient
for this procedure.
This procedure generates pretune values for correct phase-locked loop
operation.
1.
Make sure the A7 jumper/switch is in the Alter position (AU').
2.
When P-retune Adj isdisplayed, press
query to alter the correction constants
Figure 3-3. Source Pretune Correction Constants Setup
N E T W O R K A N A L Y Z E R
. Press
F R E Q U E N C Y C O U N T E R
at the